GENOSS™ DCB
Angioplasty Catheter, Balloon Dilatation, Coronary, Perfusing
GENOSS™ DCB
Description
– Paclitaxel crystal structure coating of 3μm/mm²
– Uniform coating layer with a thickness of 10μm.
– Excellent drug delivery to target lesion
More Information
Description
- Crystalline paclitaxel of 3.5 µg/mm2 coating with shellac
- Rapid deflation by large lumen diameter
- Low profile : High deliverability
Paclitaxel coated PTCA Balloon catheter
- Paclitaxel (anti-proliferative drug) delivered to target tissues through contact with vessel wall
Surface circular coating
- Homogeneous drug delivery by even coating of balloon surface
Rapidly controlled drug release
- Shellac (hydrophilic natural polymer) for rapid tissue diffusion and release of drugs
- Prevention of washout from the balloon surface upon entry into the lesion with 3-folded wing
Inhibition of restenosis
- Crystalline paclitaxel of 3 µg/mm2 coating with shellac for reduction of restenosis in the stent due to lumen loss in the late stent thrombos
Low profile delivery catheter system
- An effective system for small vessel treatment with improved deliverability
Anatomy
- Superior deliverability without compromise in complex anatomy
Hydrophilic coating
- A coating with great deliverability
Features
Drug : Paclitaxel
Drug dose : 3.00㎍/mm²
Coating thickness : Avg. 10㎛
Coating material : Shellac
Coating type : Surface circular coating
Nominal pressure : 8atm(2.00~4.00mm)
Rated burst pressure : 16atm(2.00~3.50mm), 14atm(3.75~4.00mm)
Distal shaft coating : Hydrophilic coating
Distal shaft : 2.5Fr(0.82mm)/2.8Fr(0.92mm)
Proximal shaft : 2.0Fr(0.68mm)
Lesion entry profile : 0.017”(0.43mm)
Usable length : 1450mm
Min. guiding catheter ID : 5Fr(1.42mm),0.056”
Max. guide wire OD : 0.014”(0.36mm)
Catheter type : Rapid exchange
Distal shaft coating : Hydrophilic coating
Indicant
Treatment of in-stent restenosis due to stent late lumen loss – Used for Percutaneous coronary intervention (PCI)
Crystalline Paclitaxel

Inflation / Deflation time

Folding shape & coverage


Large inflation lumen


Lower Profile

Balloon profile


Sideport profile

Serial Quantitative Coronary Angiographic Results at 6 months

Clinical Follow-Up at 6 months

Model Name




Compliance chart